Sample
Programs Downloading
Suitable for inspecting and checking iron-removing magnetic
bars in the food processing process.
Suitable for measuring residual magnetism of electronic components.
The highly reputable TM Series has moved a step forward with
new functions that support various measurement and inspection
scenes.
New model of Tesla meter equipped with an
external output function.
Features:
Digital and analog output terminals (external output terminal)
are provided.
By connecting an optional cable (Model TM-701 DTC) to an
external output terminal, digital (or analog) data can be
transmitted to a personal computer.
The probe has been reduced in width further to enable measurement
in a very limited space.
Small size and light weight.
Large and easy-to-read display.
The alternate magnetic flux density, in addition to direct
magnetic flux density, can be measured.(50/60 Hz)
A wide measuring range from 0 to 1500 mT (0 - 15,000 G).
(Direct current)
The high resolution measuring mode ensures more accurate
measurement. (Resolution 0.01 mT =0.1 G)
The unit of display can be changed over between “mT
”and “G ”.
High dustproof by use of sheet keys.
The auto power off function prevents useless waste of the
battery.
The probes can be replaced without troublesome calibration.
Application:
Measurement of residual magnetism in machined works.
Measurement of magnetic flux in magnetism-applied products.
Measurement of magnetic flux of motors.
Measurement of properties of magnetic materials.
Inspection of mechanical stress in stainless steel products.
Measurement of magnetic flux of magnets.
Inspection of natural magnetization of various steel materials.
| Model |
TM- 701 |
| Detecting Contents |
DC magnetic flux density AC magnetic flux
density
Polarities (N,S) (50/60Hz) |
| Indication Range |
mT/G selection |
1T = 10000G |
| Indication Range |
0 ~1500mT |
| Measuring Mode
Measuring Range
Analytical Ability
Indication Accuracy |
Measuring Mode |
Measuring Range |
Analytical Ability |
Indication Accuracy |
| DC × 1 * |
0 ~ 200.0mT |
0.1 mT |
±(5%of rdg.+ 3digit) |
| 200.1 ~1500.0mT |
1 mT |
±(5%of rdg.+10digit) |
| DC ×10 |
0 ~ 150.0mT |
0.01mT |
±(5%of rdg.+ 5digit) |
| AC × 1 * |
0 ~ 200.0mT |
0.1 mT |
±(5%of rdg.+20digit) |
| 200.1 ~ 750.0mT |
1 mT |
| AC ×10 |
0 ~ 75.0mT |
0.01mT |
| Function |
Zero reset |
Polarity indication |
| Max. detect value hold |
Auto power off (can be canceled) |
| Display |
LCD |
| Mode |
Detect value |
Numeric value (digital) |
| Polarity |
Alphabets (N/S) |
| Operating Temperature |
0 ~+40 °C |
| Power Source |
Battery SUM- 3 (1.5 V) x 4 pieces |
| Dimension |
Height 140mm(5.51)×width 64mm(2.51)
×thickness 30 mm(1.18) |
| Mass |
Approx.250 g/0.54 lb (including batteries
and probe) |
| Accessories |
Probe, batteries and carrying case |
| Option |
Digital output cable (TM-701DTC) |
*The measuring ranges are automatically selected.
(Note) This product does not correspond to electromagnetic
wave measurement.
Sample